Item type |
学位論文 / Thesis or Dissertation(1) |
公開日 |
2017-04-19 |
本文言語 |
|
|
言語 |
eng |
タイトル |
|
|
タイトル |
Reliability of distance measurements in dental CBCT images |
著者 |
森, 友理恵
Mori, Mori
|
キーワード |
|
|
主題Scheme |
Other |
|
主題 |
CBCT |
キーワード |
|
|
主題Scheme |
Other |
|
主題 |
Accuracy |
キーワード |
|
|
主題Scheme |
Other |
|
主題 |
Measurements |
資源タイプ |
|
|
資源タイプ識別子 |
http://purl.org/coar/resource_type/c_db06 |
|
資源タイプ |
doctoral thesis |
アクセス権 |
|
|
アクセス権 |
open access |
|
アクセス権URI |
http://purl.org/coar/access_right/c_abf2 |
抄録 |
|
|
内容記述タイプ |
Abstract |
|
内容記述 |
We evaluated the accuracy of linear measurements in dental CBCTbased images for various sites in the field of view (FOV). We used a phantom of 30 stacked commercially available CDR disks. The phantom was scanned with two conebeam computed tomogra phy (CBCT) units, a Kyocera Prevista Uni3D MultiOS and a Morita 3D Accuitomo F 17,and linear measurements were obtained with a digital caliper. The error characteristics were different for each device. There were similar errors on the edges of the FOV at 90 kV. Measurement errors in the upper and lower regions of the FOV tended to be slightly larger than in the middle. It is important to understand that the accuracy of linear meas urements in CBCT images varies depending on their location in the FOV. |
学位名 |
|
|
学位名 |
博士(歯学) |
学位授与機関 |
|
|
|
学位授与機関識別子Scheme |
kakenhi |
|
|
学位授与機関識別子 |
34408 |
|
|
学位授与機関名 |
大阪歯科大学 |
学位授与年月日 |
|
|
学位授与年月日 |
2017-03-10 |
学位授与番号 |
|
|
学位授与番号 |
甲第794号 |